Readying the Spectrograph
The Coudé Camera Room
The Coudé Camera Room contains the
spectrograph optics and detector. Observers
enter the camera room to open or close the covers of the spectrograph
optics, and to
fill the CCD's LN2 dewar. To minimize
dust and
dirt in the camera room, observers are asked to clean their shoes in the
machine pictured at right, located just outside the door to the anteroom
(see next section). Visits to the camera room should be kept to a
minimum to protect the delicate optics. Covers are usually opened at the
beginning of a run, as described below, and left open until run's end.
Note that the echelle grating, located just to the left of the door into
the camera room, is particularly delicate. Once damaged, it cannot be
repaired. Please take very special care when near the grating.
The Anteroom: CCD Controllers & Stickymats
An anteroom separates the camera room from the basement. A disposable
stickymat on the anteroom floor further protects the camera room from
dirt. Step several times on the mat before proceeding to the camera room.
If it does not feel sticky, please peel up present mat to expose a fresh
one underneath (we suggest grounding yourself to the anteroom's metal wall
when doing so to avoid a possible nasty static shock). If the worn our mat
is the last one, please notify a staff member.
NB: As of 4 Dec 2008, there is a single CCD power supply and controller
for both dewar 6 and 8. The below info is out of date.
At the time of this writing, both CCD controllers are located on a shelf
in the anteroom (plans are underway to move at least one to the camera
room, nearer the CCD). In the picture at left, the lefthand controller is
for dewar #6, the right for #8. Observers don't usually need to do anything
with the controllers, but should note the CCD temperature displayed on the
front panel of the one they are using, and check that the temperature is
constant to within a few degrees on subsequent visits to the camer room.
Opening the Spectrograph Optics
The Hamilton optics are suspended from the large I-beam on the east
side of the camera room (left when looking down into the room).
Three of them must be opened: the collimator, the cross-dispersing
prisms, and the echelle grating. They are presented below in the order in
which they should be opened. The grating is the most susceptible to permanent
damage. To afford it maximum protection, it should always be the last optic
opened and first closed.
1. The Collimator
Three collimating mirrors are suspended from the right side of the I-beam.
Only one is used with the Hamilton, but observers occasionally
confuse it with one further up the I-beam. The Hamilton collimator is the
middle one of the three, and is located adjacent to the wooden platform
about two-thirds of the way down the camera-room steps.
The collimator cover is hinged along its lower edge, and is opened
by pulling on a small handle along its upper edge. The latch is magnetic
and requires a fairly firm pull to release. When fully open, the collimator
cover hangs from its hinged edge.
Some observers use a mask on the collimator, slightly smaller than the
telescope pupil, in order eliminate any systematic effects that might arise
from the slight shift of the pupil's position on the collimator as a function
of telescope hour angle. Using the mask does result in minor light loss.
The collimator is pictured above with and without the mask installed.
Note that without the mask, the telescope pupil underfills the collimator.
2. The Cross-dispersing Prisms
The cross-dispersing prisms hang from the I-beam a few feet above, and
to the left of the collimator. They are opeed in two stages, illustrated
below.
First, open the hinged flap on the west (right) side of the cross-disperser
housing, latching it in place with the wire hook attached to the I-beam.
Don't forget this step! Observers sometimes inadvertanly
neglect to do so. The spectrograph will work, but the flap will occult
at least a third of the beam.
Second, slide the metal cover on the front side of the cross-disperser
towards you, remmoving it completely from the housing. Stow it by leaning
against the step below the cross-disperser, in such a way that it will
not cause a tripping hazard.
3. The Echelle Grating
As has already been noted, because of its vulnerability to damage (not to
mention its very high replacement cost!) the grating should
always
be the last optic opened, and the first closed, and should be treated with
an extra measure of caution. Never try to open or close the grating without
adequate light to see what you are doing.
The grating is located immediately one's left on entering the camera room.
It is tempting to take a shortcut to the dewar on the way in, or to the
camera-room door on the way out, by slipping between the
grating and the nearby wall.
Don't.
The grating cover is hinged along its lower edge, and held closed by
a latch at the center of its upper edge (the latch is circled in the
top illustration at left). You may stand either in front of or behind
the grating to open the cover, but remember not to pass by the grating
surface once it is open. To realease the latch, lift its back end
then bring the the whole latch forward and up.
Hold the cover closed while unlatching so that it does not fall open.
With one hand placed on the still-closed cover, carefully lower it
until it comes to rest on the grating mount.