UNIVERSITY OF CALIFORNIA OBSERVATORIES / LICK OBSERVATORY


PFCam Setup Checklist

Last update 19 April 2000

Topics:
Mounting PFCam on Top Ring
Floor Testing/Dry Run
Mounting on Telescope
Connections between Telescope and Control Room
Computer Configuration
Misc.
Software Startup
Telco
Misc. Trouble Shooting


Mounting PFCam on Top Ring

I. Mechanical

  1. What steps?

II. Electronic

  1. What steps?

Floor Testing/Dry Run

  1. Connections between Top Ring and Control Room What is different from when on telescope?
  2. See Computer Configuration
  3. See Misc
  4. See Software Setup

Mounting On Telescope

I. Mechanical

  1. What steps?

II. Electronic

  1. What steps?

Connections between Telescope and Control Room


Computer Configuration


Misc

  1. Remove ADC covers (both of them!)
  2. Unplug flat lamps for MOS (Flat lamps for PFCam should be 12V/5W blubs, always plugged in)
  3. Put on shroud (optional, but highly recommended if all is running well)

Software Startup

  1. log in to orthrus as pfcam
  2. start software by typing pfcam start
  3. initialize/reset/test all motors
    	A. on PFCam Gui
    		hit the reset button for each of the 4 motors
    		set Dewar focus to 3.8 (nominal focus value)
    		set TV Filter to Spinrad R
    		set Dewar Filter to any position
    		set TV Horiz Pos to Center
    	B. on MOS UI 
    		set A. Prime Focus to 3.2 (close to usual best focus value)
    		set B. ADC Mode to track
    
  4. take test exposures
    1. 1s Dark (bias levels should be about 1000 for each side of chip)
    2. 1s Normal with flat field lamps on, make sure shutter is opening (counts in 1s: B ~ 3500, V ~7K, R ~15K ,I ~50K)

Telco

  1. Start telco menu program from shane by typing telcontrol prime
  2. Set Guide Matrix to 5. PFCam (--+-)

Misc. Trouble shooting:

  1. Timing board problems - Push white button on timing board (in CCD Controller where fibers enter).
  2. CCD Controller problems - Cycle power if all else fails. Turn off 34V, 15V, then main power switch. Turn on main power switch, 15V, then 34V switches.
  3. No light to TV camera - make sure ADC covers are off and that TV Filter position is Blank or Spinrad R.
  4. No light to CCD - Make sure scr F. Obs. type is set to Normal. Make sure shutter is actually opening. Make sure ADC covers are off.
  5. Half of chip reads all zero - Make sure both AL and AR coax cables are well connected to the CCD Controller and the electronics box next to the dewar.
  6. Dewar Focus Motor won't calibrate, stay calibrated or has trouble moving - Probably needs some lubrication on the screw (WD-40 or similar product). The motor has low torque and gets stuck if there is a small amount of dirt, grit or grime. After lubrication, the galil controller will need to be power cycled to remove the errors generated by the motor failures.