U
NIVERSITY OF
C
ALIFORNIA
O
BSERVATORIES
/
L
ICK
O
BSERVATORY
Shane AO Shutdown Procedures
End of Night
|
End of Run
End of Night Shutdown Procedure
If LGS mode, call Space Command and FAA to notify laser operations are done.
If LGS mode, close lsm.tcl window(s).
Make sure all loops are open:
open()
Rotate TUB to 74.0 degrees.
Put all Field Steer motors back to their nominal, on-axis positions using fieldsteer.tcl:
Uncheck Move Telescope and Auto Close Loops
Check NGS
Click
Back On-Axis
button
Check LGS
Click
Back On Axis
button
If LGS mode, Disable WFS Focus Tracking.
Put WFS focus back to NGS focus position: 4475 (22.38mm).
Turn off uplink HV power supply (Spare 2 on Line Lamp GUI).
Turn off any calibration sources, if any are on, in saopower_ui.
End of Run Shutdown Procedure
Do the End of Night Shutdown procedure above.
Turn off MEMS Heartbeat failsafe in saopower_ui
Turn off Woofer in saopower_ui
Turn timer back on for slinelamp:
modify -s slinelamp LLPTEN=on
Exit saoWFSFocusTrack.tcl (lgs software)
Exit fieldsteer.tcl
Exit NGSfieldsteer.tcl (usually not used, so may not be up)
Exit peeko
Exit ttpeeko
Exit saocon_gui (main SAO GUI)
Exit shaneao (command line AO control)
Cntrl-C to exit the message log on real.
Exit saomot_gui
Stop motor dispatchers: ao stop saomot
Exit sharcs_fe (sharcs data taker)
Exit saowheels_gui (sharcs filter wheel GUI)
Exit sharcsdisplay (sharcs image display)
Exit scriptproc.tcl
Exit eventsounds
Log onto stealthie and stop sharcs servers (procedure is listed in the ShARCS startup document).
Disable ShARCS temperature monitoring:
modify -s saomon SH_BASICDIS="today 20:00:00" SH_COMBODIS="today 20:00:00"
(substituting an appropriate time to start up monitoring again)
Exit sharcspump_gui (sharcs pressure and temperature monitor)
Turn off power to
ShARCS filter wheel
TT filter wheel
WFS Camera
TT Camera
musher
real
Ion pump
Stealthie
Jade card
12V servo power supply
5V DC Logic power supply
24V motor power supply
24V brake power supply
Lantronixes
Galil DMC 40x0 motor controllers.
Exit saopower_ui.
Disconnect ethernet, power, and laser uplink cables from SAO.
Remove SAO from telescope.
Last modified: Thu Feb 5 04:37:23 PST 2015