Observing Recipes
Darks and Biases
Focus
Flat Fields
Spectroscopic Calibration - arc lamps and sky lines
Photometric Standard Stars
Spectroscopic Standard Stars
Narrow Band Imaging
Deep Imaging
Using the Grisms
Polarimetry
Spectropolarimetry
Drift Scanning
Various observing "recipes" are described below. These are not meant to be strict rules of observing, but rather suggestions for using the tools provided by the Gemini system. Observers will, of course, have their own preferences and needs based on their science programs.
The most general approach used is to keep the star or object "on the frame" whenever possible and "dither" the telescope pointing between integrations. This technique eliminates bad pixels, and provides good flat-field response and sky subtraction and improves sampling. Obviously, for extended objects this may not be ideal.
Darks and Biases
Dark and bias frames should be taken at the beginning and end of each night.
The recipes below are for single sampling mode and very useful for
checking the noise characteristics of the chip and to make sure the
camera is in a steady state condition. If observing with a different
sampling mode, you may want to change the recipe accordingly (the
Gemini software will tell you the minimum Itime if what you set too
short).
BIAS: Press F1, enter object name BIAS, set Itime =
0.07, Coadds = 1, select SampMode 1 (single sampling) and set the
filter to Blank (B). Press F4 to take and image in
both channels.
Typical ranges in single-sampled bias frames are:
The dynamic range is about 6,000 DN for the SBRC at 24 electrons/DN giving
144,000 electrons full-well.
The dynamic range is about 8,000 DN for the NICMOS at 20 electrons/DN giving
160,000 electrons full-well.
Repeat this test in any sampling mode you plan to use. There should
be very little bias structure remaining in any of the CDS modes.
DARK CURRENT: Enter the Next Observing Setup and select Blank from the filter
list and change the Itime to 100.0 sec with 1 co-add. This setup will provide
a DARK frame.
Examine the displayed STATISTICS for the MEAN value and divide by 100 to get
DN/sec. In a steady state condition the nominal values are:
NICMOS: 0.5 DN/sec = 10 electrons/sec
NOTE: These figures are usually dominated by the
"hottest" pixels and the MODE is generally a bit smaller.
Dark current can be scaled with time (if bias structure is subtracted away).
However, it is also useful to take dark frames with exactly the exposure time
with which observations were made.
Focus
At the beginning of each night, it is necessary to optimize the focus of the
telescope with the Gemini camera. Gemini is approximately parfocal with the
Lick guide camera, but not exactly. To optimize the focus, you may choose
either a relatively bright star, for example and Elias IR standard, or a
fainter star, for example K = 11. In either case, integration time and co-adds
must be chosen to be short enough not to saturate on the star, and long enough
to average short timescale changes in the seeing. Focusing is best done with
differences of mode 1 frames (since saturation is easy to determine), but this
is time consuming. Most observers prefer to use CDS frames. Be aware, however,
that a saturated star can resemble and out of focus star!
Telescope focus may be changed using the telescope operator's POCO GUI.
Focus encoder values are displayed to the telescope operator (TO), so you will
have to ask the TO for the current focus value and the change the focus.
Flat Fields
Dome flats often seem to work quite well with bright objects, so it is
recommended to get
a complete set of dome flats in all filters; daily dome flats should be
obtained, although there is no real evidence of drift.
To take dome flats, ask the telescope operator to open the mirror
cover (and close the dome) and to put the diagonal mirror in position
3. There are three dome flat lights that may be used. The "red" lamp
is faint and suitable for direct imaging flat fields. The "blue" lamp
is brighter and useful for narrow band filters and grism flats. The
"super blue" lamp has a rheostat voltage control and can be varied in
brightness for any flat fields. When taking dome flats in 2 micron or
longer filters, it is necessary to take frames with both the lamp on
and the lamp off (dome dark) in order to remove thermal
background. This is especially important for K grism flats.
Sky flats should be obtained by dithering the telescope between
exposures. A sequence of at least 5 positions is recommended for median
filtering, with 7 to 9 being even better.
Twilight flats are also possible, though they are somewhat tricky to obtain, as
they are at any telescope. Beginning with J and proceeding through H and K and
narrow band filters is the best strategy at the end of the night.
Spectroscopic Calibration
- arc lamps and sky lines
Grism observations require relative and absolute wavelength
calibration. Relative calibration is done using the OH sky emission
lines. These lines are plentiful throughout the J, H, and K
windows. There are fewer OH lines at the long end of the K window, but
fortuitously, there are other sky emission lines which have been
tabulated which fill this region; these lines have been observed to be
repeatable over the course of the year. OH and other sky emission
lines are tabulated in:
Maillard, J.P. & Chauville, J. 1976, Journal of Molecular
Spectroscopy 63, 120
Absolute wavelength calibration is achieved using noble gas arc
lamps. The Gemini equipment included the following lamps: Neon, Argon,
Krypton, Hydrogen, Hg-Vapor, and CO2. These lamps
are mounted on the inside of the telescope, on the side toward the
control room door. Only one may be mounted at a time. The default installed
lamp is Argon, if you require a different lamp contact a support astronomer
in advance of your run to be sure the lamp is available.
The lamps must
be changed manually using the lift. The lamp is plugged into the first outlet
of the lamp controller, Spare1_(110SSR). It is
turned on by clicking the Spare1 button in
slinelamp_fe. Also be sure that the
diagonal mirror is in position 2 when doing arc lamp exposures.
The lines from each lamp are tabulated in Tables of
Spectral Lines of Neutral & Ionized Atoms, Stringov, A.R. &
Sventitskii, N.S., IFI/Plenum Data Corp, 1968
Argon line identifications for Gemini grisms are
now available listing approximate pixel position with the wavelength of the line for each
grism.
Sky lines identifications are available at http://people.bu.edu/clemens/mimir/night_sky_spectra.htm.
The usual method of observing standard stars is to use a script to move the
standard star to two or more various positions on the chip.
If seventh magnitude Elias standards are used with broad band filters, under
good seeing conditions, it is essential to use the Single Sampling mode with
the shortest allowed integration time. Even so, those brighter than K = 7.0
will saturate.
For the fainter (9 to 13th magnitude) UKIRT standards, the CDS mode can be
used, but care should be taken not to saturate on the brightest ones, and not
to underestimate the required exposure time for good signal-to-noise ratio on
the fainter stars (about 5 minutes).
IR standard star lists are available
both on-line and in the 120" control room.
This section is not intended to be a complete reference for spectroscopic
standard stars. Most grism spectroscopy programs will choose appropriate
standard stars in advance of arriving at the telescope. This section is meant
simply as an introduction so that "spur of the moment" spectroscopy can be
done if it is required.
The purpose of a spectroscopic standard is to provide a calibration for the
effect of the atmosphere on the spectrum from an object. The "perfect"
standard, would therefore be one with a featureless spectrum. The closest
approximation to this in the astronomical world is to choose stars whose
spectra have few features in the waveband of interest. Once a stellar type
is identified, appropriate stars can be found in the SAO catalog or the Bright
Star catalog. There are copies of both in the 120" control room. When choosing
standards, it is important to avoid variable, peculiar, binary, and otherwise
"unusual" stars.
It is important to take spectroscopic standards at similar airmass and close
in time to the objects they will be calibrating. As a rule, one should seek
standard stars within 0.2 airmass and 30 minutes of the observation.
J band: Late F and early G giant stars have few features in the J band. They
are plentiful in most regions of the sky.
H band: O stars are the best standards for the H band. Unfortunately, they are
rare. A list of all the O stars in the Bright Star Catalog is below.
K band: G and K main sequence stars are featureless in the K window except for
Br-gamma. If the 2.16 micron region is not of interest, these stars will
suffice. If that region is necessary, the G or K star spectrum can be
interpolated over (using a model atmosphere is possible). Alternately, a
separate calibration can be taken for that part of the spectrum. Early to
middle A stars have a feature-filled spectrum except in the region Br-gamma.
Gemini has available more filters than can be installed at one time. While
filters are not routinely changed, and it may be a year or more before the
current setup is altered, this section will detail all the filters available.
The User Interface Software will always have and accurate filter list.
This section is more of a data sheet than a recipe as the procedure
you will use ultimately depends on your program. The resolution numbers were
measured from the manufacturer's traces, so they will be accurate to within
a few percent. The other figures are much more variable, by perhaps as much
as a factor of 2, due to the intrinsic variability of the backgrounds and sky
transmissions. In addition, the observed transmission of filters in the K
window will vary depending on which dichroic the light passes through; there
may be as much as a 20% difference.
Most of the narrow band filters exhibit "ghost" images; these are internal
reflections off the filter and/or subsequent optics. The strength of these
ghost images varies from filter to filter, but is typically on the order of
2% and always less than 5% of the signal from the "ghosted" object. Some
"ghosts" move anti-parallel to the object when the telescope is dithered.
Others move with the object and are thus harder to distinguish from real
objects. It is very important to consider the possibility of ghosts in
narrow band data. It is important to "dither" observations with narrow band
filters; ghosts may then be removed in data reduction through appropriate
combinations of pixel rejection and pixel masking.
Often narrow band observations will require very long integration times to
achieve desired signal to noise ratios. Typically, one desires to be background
limited; that is to have the shot noise in the background signal be several
times the read noise. This calculation can be done based on the background
levels provided in Filters
and the readnoise estimates provided for each sampling
mode in Sampling Modes.
Many narrow band filters also exhibit interference fringes due to reflections
internal to the detector; these effects are reasonably stable and should
flat-field or sky subtract away.
This section outlines the suggested method for performing deep near-infrared
imaging with the Gemini camera on the Lick Observatory 3-m telescope. We have
successful results using a dithering technique in which the total integration
time is divided into numerous shorter integrations, displacing the telescope
in a regular pattern between integrations. This way, the disregistered frames
can be median filtered together to build a master "sky frame" which is used
to flat-field the individual frames. Then the frames are registered and
median filtered together to obtain a single deep image. Similarly, you
may use the "sky frame" to subtract the sky background and use a dome or
twilight flat for flat-fielding. Some observers prefer one method over the
other.
J, H, and K images may be obtained simultaneously in some programs, and
each can reach comparable limiting magnitude by integrating J for 1/3 of the
K integration and in H for the remaining 2/3 time. Overhead is greatly reduced
by writing a script which will automate the telescope moves, filter changes,
and integration times. Examples of scripts are found in the
c:\gemini\script directory of the Gemini 486 PC.
If you know what resides in your field, it is simple to calculate the
necessary parameters in this observing strategy. However, if the purpose of
the deep imaging is to find out what is there, reasonable
guestimates are required. The parameters are outlined below.
Integration times. Deep imaging is broken down into three different
integration times: time per co-add, time per frame, and total integration
time. Of these, the time per frame is the only one not normally encountered
with infrared array observing.
The time per co-add must be as large as possible without the brightest
interesting objects (usually the background level) reaching the
non-linear part of the wells in the array. This also ensures that the
frames are background-limited. Total integration time depends on how
deep the observer wishes to reach. Time per frame is a bit more
tricky. If is important to have a large number of frames taken close
in time for purposed of constructing a good sky frame. Optimally, we
tend to use frame times of about 2 minutes; any longer than this, and
changes in the sky may prevent construction of an accurate
flat-field. In addition, if objects in the field get too bright in the
individual frames, a "shadowing" effect may appear in the final image
which surrounds bright objects with inverse images of themselves.
Dither pattern. Once the total integration time and time per frame
are determined, the required number of individual frames is set. A dither
pattern should be constructed that is centered on the field of interest
and symmetric around it, so that there is sufficient overlap of frames in all
directions. We tend to use a square 3 x 3 pattern of nine frames.
Telescope displacement. The greater the telescope move, the smaller the
deep central overlapping region becomes. However, the displacement must be
great enough for extended objects to "clear" themselves, such that they will
not fall onto common pixels from frame to frame. A displacement of 14" is a
good default value.
Once the observing parameters have been chosen, some final issues must be
considered.
Observing procedures for the grisms are still being optimized, but our
experience in the July 1994 run was very efficient and fruitful. This
section will not address the necessary preparation which precedes the
actual observing, such as finding suitable standard stars and
calculating the required exposure times to obtain the desired
signal-to-noise ratios; however, these tasks are usually more
important for spectroscopy than the analogous tasks are for other
types of observing. For more information about spectroscopic observing
and data reduction, please see Don Figer's 1995 Ph.D. thesis and the
IRAF documentation on long slit spectra.
The following steps, taken from experience in the July 1994 run, are a
good first start for most spectroscopy programs. Before beginning
observations, check which column the slit is centered on (nominally
128, but it may vary slightly depending on mechanical alignment). For
each observation, you will need to decide a "start spot" along the
slit to place the object; which row this spot is on depends on how
many nod positions you plan to use. You will need to guide during the
observations. If you are only using two nod positions, ask the
telescope operator about using the alternating guide box positions
("A" and "B" on the guider).
Note: To get Arclamps
Polarimetry
The Gemini polarimetry mode allows polarimetry data to be taken with a
minimum of observer involvement after the initial setup phase. To
access this mode, go into "General Setup" under the "Setup" menu item;
then select the "polarimetry" option under "observing mode". This sets
up the software for polarimetric observations. The software is now
configured so that the waveplate and polarizer are automatically
placed into the beam. Check that this is so by looking at the display
parameters in the upper part of the screen. Assuming all is correct,
you are ready to begin observations.
There are two basic differences when observing in polarimetry mode
vs. imaging mode. The first is the software now automatically takes 4
images each time the "GO" command is issued. These are taken in the
order of 0, 45, 22.5, 67.5 degree positions of the waveplate. The
second, more visible change, is in the TV display area.
On the display, all four waveplate position images are shown in
smaller versions from the normal display. Each frame is displayed as
the data is received. Also, there are 4 new command functions,
F1 thru F4. These allow you to
select and individual image with which to interact on the TV
display. Setting the offset and gain for one image will set it for all
4, though you must manually go to each image (via the
F1 thru F4 commands) for the screen
to redisplay. [Bug: it doesn't always automatically display when the
new image is selected. Easiest way around this is to press
PgUp followed by PgDn which will
restore the original stretch set up on another image.] All other
commands are the same.
Test frames are still available in the polarimetry mode and they are a
single frame. This image is displayed in the upper left image spot and
is generally taken in the waveplate angle position of 0.
To improve the duty cycle, polarimetry mode will begin the next
integration in the series of four before saving the first frame to
disk. If the exposure time is short enough that the next frame arrives
before the first is saved the software will beep while it finishes
saving the first integration. In most cases (exposures longer than
about 10 seconds), this will not be a problem.
The observing setup for the next set of polarimetric observations
(four frames) may be input at any time during an observation. The
changes will not be made until the current suite of four positions is
finished.
Warning: Exposure setups with exposures less
than 1.5 seconds will fail in polarimetry mode, requiring a reboot of
the computer. If you need to do short exposure polarimetry, you will
have to do it in Imaging mode, put in the Polarizer, Waveplate, and
manually rotate the waveplate through the four positions.
Spectopolarimetry
This mode works analogously to the polarimetry mode. Selecting
"SpecPol" in the "General Setup" menu moves the waveplate, the calcite
polarizer and the double slit aperture into position. The image
display is again split into four images.
Drift Scanning
Since we are limited to a maximum drift rate of 1"/sec, that number will be
fixed and just the declination comes into play:
cos(51degrees) * 1"/sec = 0.629"/sec
The Gemini pixel scale is 0.7"/pix, thus if we want the time per pixel we
need to divide the rate by the pixel scale:
0.7"/pix / 0.629"/sec = 1.11sec/pix
This number does not take into account the overhead of reading out the array
which cuts the sec/pix by a factor of ~2. So the maximum possible integration
time is 0.55sec. This should be adjusted by a standard star or other point
source at the same Dec as the object that is being imaged. Then the number
can be split into an actual exposure time and co-adds.
Detector A starts at pixel 129 and next 16 (to 144)
Drift scanning is like single sampling. Typical dark/bias levels are:
A: -4,500 in a single co-add.
In drift scanning you get 16 columns worth of these values. Therefore typical
dark/bias levels are:
A: -4,500 x 16 = -72,000
You can plot these best by using the following gains and offset settings:
A: Gain = -0.0125 Offset = -65,000
Both plots show horizontal "banding" structure. F10 (spPlot)
can give a scan (inverted) along any line to show the ramp up at both sides.
If multireads = 128, the "right hand" half of each display is shown.
Saturation: Drift Scanning is like single sampling x 16.
Therefore if bias is:
A: -72,000 then the saturation is ~+500 x 16 = 8,000
Note: These saturation levels are really 2/3 full well:
L exposures must be less than 0.010sec (good clear skies) and probably more
like 0.007sec.
SBRC: 2.0 DN/sec = 50 electrons/sec
It may be necessary to refocus the telescope during the night. It is worthwhile
to use the profile function periodically to check the focus.
Oliva, E. & Origlia, L. 1992, A&A 254, 466
Ramsay, S.K., Mountain, C.M., and Geballe, T.R. 1992 MNRAS 259, 751
Bright Star Catalog O Stars HR Name HD RA (2000) Dec (2000)
V B-V U-B R-I Sp. Type 65 --- 1337 00 17 43.0 51 25 59
6.14 -0.13 -0.97 0.00 O9IIInn 1209 --- 24534 03 55 23.0 31 02 45
6.10 0.29 -0.82 0.00 O9.5ep 1228 46Xi Per 24912 03 58 57.9
35 47 28 4.04 0.01 -0.92 -0.01
O7.5III 1542 9Alp Cam 30614 04 54 03.0
66 20 34 4.29 0.03 -0.88 0.00
O9.5Ia 1712 --- 34078 05 16 18.2
34 18 43 5.96 0.22 -0.70 0.00
O9.5V 1852 34Del Ori 36486 05 32 00.4
00 17 57 2.23 -0.22 -1.05 -0.22
O9.5II 1879 39Lam Ori 36861 05 35 08.3
09 56 03 3.54 -0.18 -1.03 -0.17
O8III((f)) 2422 --- 47129 06 37 24.1
06 08 07 6.06 0.05 -0.88 0.00
O8p 2442 --- 47432 06 38 38.1
01 36 49 6.21 0.15 -0.82 0.00
O9.5II 2456 15 Mon 47839 06 40 58.7
09 53 44 4.66 -0.25 -1.07 -0.22
O7V((f)) 2467 --- 48099 06 41 59.3
06 20 42 6.37 -0.05 -0.94 -0.05
O6.5V 7574 9 Sge 188001 19 52 21.8
18 40 19 6.23 0.01 -0.92 0.00
O7.5Iaf 7589 --- 188209 19 51 59.1
47 01 39 5.62 -0.07 -0.97 -0.12
O9.5Ia 7767 --- 193322 20 18 07.0
40 43 56 5.84 0.10 -0.78 0.01
O9V 8023 --- 199579 20 56 34.7
44 55 30 5.96 0.05 -0.85 0.00
O6V((f)) 8154 68 Cyg 203064 21 18 27.2
43 56 45 5.00 -0.01 -0.94 0.00
O7.5III:n 8281 --- 206267 21 38 57.6
57 29 21 5.62 0.21 -0.74 0.00
O6.5V((f)) 8327 --- 207198 21 44 53.3
62 27 38 5.95 0.31 -0.64 0.17
O9Ib-II 8406 14 Cep 209481 22 02 04.6
58 00 02 5.56 0.06 -0.86 0.00
O9Vn 8428 19 Cep 209975 22 05 08.9
62 16 48 5.11 0.08 -0.84 0.03
O9.5Ib 8469 22Lam Cep 210839 22 11 30.7
59 24 52 5.04 0.25 -0.74 0.15
O6I(n)fp 8662 10 Lac 214680 22 39 15.7
39 03 01 4.88 -0.20 -1.04 -0.22
O9V
Arc light is reflected into the instrument from white card on the back of the
acquisition mirror.
Column # to Wavelength Conversions for some
Popular Lines Band Column # Wavelength
Line H 59 1.738 Br-10 H 83 1.700 HeI H 90 1.688 FeII H 93 1.684 Br-11 H 118 1.644 FeII H 119 1.643 Br-12 H 138 1.613 Br-13 H 153 1.589 Br-14 K 61 2.290 CO K 124 2.166 Br-gamma K 147 2.121
H2 K 152 2.113 HeI K 169 2.078 CIV K 180 2.058 HeI
Detector B starts at pixel 113 and next 16 (to 128)
B: -7,200 in a single co-add.
B: -7,200 x 16 = -115,200
B: Gain = -0.0625 Offset = -115,000
B: -115,200 then the saturation is ~-3,200 x 16 = -51,200
Max for A: ~+1-,400 DN
Max for B: ~-35,800 DN
Last modified: Wed Oct 20 14:20:26 PDT 2010
by Elinor Gates